Chapter 4 Advanced
◎
CAL – AVERAGE
This correction reduces noise content from the image signals by signal averaging of the thermal image,
improving the noise equivalent temperature difference (NETD).
(1) On the CAL screen, select “IMAGE
QUAL” and press CENTER.
(2) Select “AVERAGE”, and press CENTER.
(2) Select Low or MED or High, and press CENTER.
Notes
- This setting is enabled while in the run state.
- The status icon changes to (Gray) for low sensitivity, to (Green) for middle sensitivity, or to
(Blue) for high sensitivity.
- With the higher averaging, the image responds to quicker movement and shows residual image.
95691-2935-0000
4-39
Summary of Contents for Thermo Gear G100