CAL – REF CAL
This item is used to execute the reflection correction (Ref-Cal).
How to operate
1. “EXEC CAL. IS IT OK?” is displayed on the screen. Select “YES” with NEAR(Left) or
FAR(Right), and then press CENTER.
2. When “PUT THE LENS CAP.” is displayed, attach the lens cap to the camera.
3. Select “YES” with NEAR(Left) or FAR(Right), and then press CENTER.
4. “REF CAL” is displayed on the screen. Please wait until the correction process is completed (do not
remove the lens cap).
5. When the message disappears, it indicates a completion of the correction. After that, remove the lens
cap.
Notes
•
Reflection correction is unavailable when the system is in the freeze status.
•
The lens cap should be adapted to ambient temperature sufficiently before attaching it to the camera.
•
When the range is changed or the power is turned off, reflection correction is disabled. In this case,
execute it once again.