Labfors 5 – Version for Microorganisms
Cleaning and Maintenance
22 January 2019
Page
173
of
205
10.10 Cleaning the Heater Pad
DANGER
Damp cleaning of a heater pad, which is connected to energised
equipment, can be perilous.
– Always disconnect the heater pad from the basic unit be-
fore cleaning.
– Preferably clean the heater pad dry.
– Always let the heater pad completely dry after cleaning with
a damp cloth and before next use.
If required, clean the heater pad as follows:
Procedure
1.
Unplug the heater pad from the basic unit.
2.
Wipe the heater pad with a dry cloth.
If necessary, use a damp cloth (water only!).
ATTENTION
Cleaning the heater pad with soap or aggressive detergents
may damage the silicone sheathing.
3.
Let the heater pad completely dry after cleaning with a damp
cloth.
4.
Check the heater pad and cable for damages (cracks, kinks
etc.), replace as necessary.
10.11 Maintenance Plan
WARNING
Non-compliance of this maintenance plan contains a high risk!
It is the responsibility of the user, that this maintenance plan is
complied with. Non-compliance will lead to exclusion of liability
(see General Terms and Conditions).
The required maintenance for reliable operation is described in the
following chapters.
Reduce the maintenance intervals in case increased abrasion is
detected during regular checks.