Labfors 5 – Version for Microorganisms
Cleaning and Maintenance
22 January 2019
Page
163
of
205
INFORMATION
This step is only necessary for the type of clamping adapter for
10 mm ports. All other clamping adapters can be directly un-
screwed and pulled out from the port or threaded adapter in the
port by hand.
2.
Carefully pull the clamping adaptor together with the sensor
out of the port / threaded adapter in the port.
3.
Ensure the O-ring of the clamping adapter does not get lost.
As required:
4.
Depending on the type of clamping adapter, loosen the slotted
screw or the hollow screw and carefully pull the sensor out of
the clamping adapter by hand.
5.
Ensure that the insulation on the sensor does not get dam-
aged.
10.3.3 Removing Hoses, Filters and Pump Heads
To later clean reagent hoses and pump heads, they must be re-
moved from the reagent bottles and from components of the cul-
ture vessel.
INFORMATION
To avoid damage, never dismantle the pump heads. Always re-
place a damaged pump head along with the pump hose, and
vice versa.
Proceed as follows:
Procedure
1.
Remove cable ties (e.g. with a side cutter) so that the hoses
are not damaged.
2.
Pull hoses off the culture vessel and the reagent bottles.
3.
Remove and dispose of filters for pressure equalisation and
hoses from reagent bottles.
4.
Ensure that the inlet air filter is clean, dry and not blocked. If
this is not the case, dispose of it.
INFORMATION
If the filter for pressure equalisation and the corresponding
hoses have been used several times, ensure that the filters are
always dry and clean.