Labfors 5 – Version for Microorganisms
Cleaning and Maintenance
Page
162
of
205
22 January 2019
10.3 Removing the Vessel Top Plate and Accessories
All accessories must be removed for thorough cleaning of the indi-
vidual parts of the culture vessel. This is described in the following
sections. The cleaning itself is described in the chapter “Cleaning
and Storing Individual Parts”.
The cleaning of the hoses with pump heads, the basic unit, operat-
ing panel and the exit gas cooler are described in separate sec-
tions.
Sensors from third-party manufacturers are cleaned according to
their manufacturer’s specifications.
10.3.1 Removing the Exit Gas Cooler
Proceed as follows:
Procedure
1.
Unscrew the exit gas cooler from the vessel top plate port by
hand.
Ensure that the O-ring does not get lost.
2.
Gently release the hose clamp with the hand wheel, pull off
the exit gas filter and dispose of it.
3.
Remove the pressure hose piece to thoroughly clean the exit
gas cooler. (For details see the chapter “Cleaning the Exit Gas
Cooler”.)
10.3.2 Removing the Sensors
Sensors are either mounted directly into the ports or by means of
an electrode holder or clamping adapter. For removal of the sen-
sors, proceed as follows:
Sensors without holder/clamping device
Procedure
1.
Carefully unscrew the sensor by hand (no tools!) from the ves-
sel top plate port.
Sensor with electrode holder
Procedure
1.
Carefully loosen and unscrew the sensor on the hollow screw
of the electrode holder from the port by hand.
2.
Carefully unscrew the sensor from the guide bar and gently
pull it out of the electrode holder.
Sensor with clamping adapter (antifoam and level sensor)
Procedure
1.
Loosen both slotted screws next to the sensor in the port.