Labfors 5 – Version for Microorganisms
Accessories
Page
90
of
205
22 January 2019
5.24 Sterile Filters
Sterile filters are used to protect against contamination in both the
gassing line and the exit gas line. In addition to this, all reagent
bottles used for pressure equalisation must be fitted with an open,
short section of hose with a filter.
All the sterile filters in the scope of supply are autoclavable, dispos-
able filters with PTFE diaphragms.
INFORMATION
Sterile filters must be clean and dry at all times, and should thus
ideally be replaced after each use.
Ø 37 mm, marked red
Application
Supply air
Retention rate
0.2 µm
Ø 37 mm, marked green
Application
Exit gas
Retention rate
0.3 µm dry
1.0 µm wet
Ø 25 mm, not marked
Application
Super Safe Sampler
Retention rate
0.2 µm
Diaphragm
PTFE