Labfors 5 – Version for Microorganisms
Options
22 January 2019
Page
55
of
205
All information about the ABER Futura system is available in the
separate documentation provided by the manufacturer.
4.8 Redox Measurement
Measures the reduction/oxidation potential (redox) in the medium.
4.8.1 Measuring Systems (Sensors)
Two systems are available for redox measurement:
METTLER version (analogue)
Classic combined sensor (oxidation reduction potential meas-
urement against a reference)
Type: 405-DPAS-SC-K8S
Manufacturer: METTLER TOLEDO
Measures the reduced potential in the medium in the range
from
-2,000 mV to +2,000 mV.
To use the sensor, the basic unit must be equipped with a corre-
sponding connection.
HAMILTON version (digital)
Classic combined sensor (oxidation reduction potential meas-
urement against a reference) with integrated electronics
Type: Easyferm Plus ORP ARC
Manufacturer: HAMILTON
Measures the reduced potential in the medium in the range
from
-1,500 mV to +1,500 mV.
If the basic unit is configured for Hamilton sensors, the redox sen-
sor can be connected instead of the pO
2
sensor. If the sensor is
configured in addition to the Hamilton pO
2
sensor or if the basic
unit is configured for Mettler sensors, an additional connecting ca-
ble is required.
4.8.2 Calibration
The redox sensor is usually not calibrated/adjusted. However, the
calibration of the sensor can be checked using the appropriate re-
dox buffer solution.