Labfors 5 – Version for Microorganisms
Options
Page
54
of
205
22 January 2019
4.6.6 Specifications
Description
Value
Sensor type and
manufacturer
ASD12-N
Available path
lengths
OPL01: for very high cell densities
OPL05: for higher cell densities
OPL10: for lower cell densities
Material sensor
316L, electropolished and sapphire glass
Measuring principle 1-channel absorption of light
Measurement wave-
length
840 nm – 910 nm
Detector
Silicone photodiode (hermetically sealed)
Measuring range
0 – 4 CU (concentration units)
Protection
IP 68
Light source
Hybride LED
Temperature
Shutdown at
50 °C
1
1)
The measuring system automatically switches off, if the temperature
of the sensor in the medium exceeds 50 °C during operation (thermal
shutdown). Once the medium has cooled down, measurement auto-
matically continues.
4.7 Permissive Measurement
The sensor with the corresponding transmitters must be purchased
directly from the manufacturer ABER. INFORS HT offers a connec-
tion to the transmitter on the basic unit.
Calibration is performed according to the manufacturer’s guidelines
directly on the transmitter.
Manufacturer: ABER
System: Futura
Measured parameters:
Permittivity in the 0 to 400 pF cm
-1
range
and:
Conductivity in the 0 to 40 mS cm
-1
range
Sensors of the ABER Futura systems measure the permittivity
(also:
capacitance
) and conductivity of the culture. This measured
data can be used to determine a correlation with the live biomass
concentration, for example, using the soft sensor in eve
®
or data
evaluation.