14 | Tune
INFICON
276 / 436
074-471-P1E HAPSITE ER Operating Manual
14 Tune
14.1 Introduction to AutoTune and Manual Tune
The tune of a Mass Spectrometer (MS) determines the quality of the mass spectrum
produced by the system. The MS performance will run an autotune upon start-up and
after 12 hours of continuous operation. Tuning is normally accomplished by the
AutoTune program, which automatically sets and adjust all parameters, however, the
user can set the parameters by manual tuning.
HAPSITE ER uses two gas internal standards which contain mass fragments that
span the mass range of interest. The internal standards are:
• 1,3,5-Tris (trifluoromethyl) benzene
• Bromopentafluorobenzene
14.2 AutoTune
AutoTune can be started from the front panel or from the laptop.
14.2.1 Starting AutoTune from the Manual Tune Screen or
Laptop Computer
1
Double-click on the
ER IQ
icon.
2
Double-click on the
Tune
icon.
Manual Tune
is an advanced user function. Refer to Set Access Level [
168] for
instructions on changing access levels.
3
Wait until the
EM and Emission
buttons on the
Control Panel
turn green. Click
the
Tune
icon.
Summary of Contents for HAPSITE ER
Page 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Page 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Page 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Page 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Page 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Page 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Page 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Page 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Page 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Page 435: ......
Page 436: ......