17 | Maintenance
INFICON
414 / 436
074-471-P1E HAPSITE ER Operating Manual
4
If the bakeout is complete, run a blank method. If the bakeout was successful
HAPSITE ER will be operational. If the bakeout was unsuccessful, a high
pressure error will occur.
17.6.1 Reactivating the NEG Pump
Reactivating the NEG pump requires having the Service Module attached to HAPSITE
ER for at least 22 hours. This procedure is the same procedure that is used to activate
a new NEG pump.
1
Click on the
Service Module
tab of the
Properties
menu.
2
Click on the
Service
button.