074-471-P1E HAPSITE ER Operating Manual
283 / 436
14.4 Performing Manual Tune
Manual tunes are a standard routine maintenance practice for HAPSITE ER and
should be performed once every 4 to 6 months to maintain optimal performance of the
unit’s mass spectrometer.
Routine manual tunes will improve the accuracy of compound identifications, and
provide indications of the health of the mass spectrometer.
The following procedure outlines the steps in manually tuning HAPSITE ER. If
performing this task for the first time, we recommend contacting INFICON directly for
support.
14.4.1 Manual Tune Variables
The goal of Manual Tune is to adjust the inputs, such that the outputs fall within
appropriate ranges.
Instrument Outputs:
• Base Peak Gain(BPG)
• Ion Percentages
• Status Column
Primary User Inputs/
Editables
• Ion Resolutions
• Ion Energies
• EM Voltage
Summary of Contents for HAPSITE ER
Page 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Page 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Page 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Page 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Page 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Page 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Page 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Page 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Page 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Page 435: ......
Page 436: ......