14 | Tune
INFICON
294 / 436
074-471-P1E HAPSITE ER Operating Manual
Emission Current
Optimizes the ionization efficiency of the ionizer.
Emission Current
is set
to achieve maximum intensity for mass 117. Range is 300 to 400. (350 is
typical.)
Extractor Volt
Optimizes the ionization efficiency of the ionizer. The
Extractor Volt
setting must be set to achieve maximum intensity for mass 117. Range is
70 to 90.
Focus Volt
Optimizes the ionization efficiency of the ionizer. The
Focus Volt
setting
must be set to achieve maximum intensity for mass 117. Range is 2 to 9.
Threshold
Threshold
determines if a measured point is used in the peak area
integration. If the point is used, the baseline is subtracted before use. The
threshold should be set within one standard deviation of the baseline.
Baseline
The
Baseline
is the mean value of the measured noise level.
Reverse rod polarity
Changes the rod polarity on the mass filter and select the rod polarity that
provides optimal performance at mass 117.
TP
The total MS pressure. Must be below 6E-03 for instrument to operate.
Running tune base
peak gain
Current measured Base Peak Gain (BPG).
Note:
The Base Peak Gain (BPG) will switch to red when BPG is outside
the target range.
Auto resolve
Adjusts the resolution of all mass peaks to the target resolution.
Save tune
Save the tune file.
Mass calibration
Verifies and corrects the ten calibration masses for correct location within
the mass range.
Full scan
Switches between full scan display mode and peak scan display mode
Short AutoTune
Starts the AutoTune function
Noise check
Scans a “quiet” mass range (no peaks) to determine the electronic baseline
and threshold noise level.
Tune checkup
Runs a mass calibration and noise check.
Zoom
Enables the cursor to zoom into full scan or a section of the screen.
Mass adjust
Enables the cursor to select and drag a mass peak to a different position
on the mass axis.
14.4.11 Peak Scan Window
The
Peak Scan Window
, as shown below, can be used to manually tune the mass
peak.
Summary of Contents for HAPSITE ER
Page 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Page 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Page 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Page 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Page 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Page 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Page 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Page 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Page 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Page 435: ......
Page 436: ......