INFICON
Operate the Instrument in Portable Mode | 9
074-471-P1E HAPSITE ER Operating Manual
59 / 436
If the system is preparing a SURVEY run and an ANALYZE method is
desired, touch the PREPARE ANALYZE button. Likewise, if an ANALYZE
method is being prepared and a SURVEY is desired, touch the PREPARE
SURVEY button.
9.1.1 Emergency Mode (EMER MODE)
In an emergency, the concentrator cleanout can be bypassed to allow for faster
startup. This is not recommended for everyday use. While Emergency Mode is active,
the concentrator cleanout will continue to be skipped until Emergency Mode is exited.
To place the system into Emergency Mode:
1
Touch
EMER MODE
while the
PREPARING SYSTEM
message is displayed.