21 Appendix B: Calibrating Gas Mixtures
21.1 Acquisition, Preparation, and Handling
CAUTION
Failure to calibrate the instrument may give you inaccurate
identifications when sampling.
WARNING
When using chemicals, wear the appropriate PPE according to SDS.
HAPSITE ER (or any GC/MS instrument) must be calibrated at one or more
concentration levels of the organic compound(s) of interest for quantitative analysis. In
the case of the HAPSITE ER, the compounds of interest must be supplied to the
instrument as a gaseous mixture of known volume/volume composition (mole/mole %
or ppmv levels in air or nitrogen) and at atmospheric pressure.
There are a number of important factors to consider in acquiring, preparing, and
handling gaseous standard calibration mixtures. These can be organized in three
groups:
1. How to establish the desired concentrations of the required compounds. See How
to Establish the Desired Concentrations [
426].
2. Correct delivery of the mix to the inlet of the HAPSITE ER. See Correct Delivery of
the Mix to the Inlet of the HAPSITE [
3. Gas cylinder safety, contamination checks, and corrective steps in equipment. See
Gas Cylinder Safety, Contamination Checks, and Corrective Steps [
430].
21.1.1 How to Establish the Desired Concentrations
There are two basic ways to obtain several concentrations of a given mix of
compounds. The compounds can be bought, premixed to specification, in cylinders
containing the several concentrations desired. A master cylinder of the compounds
can also be purchased at the highest concentration needed, and diluted to the lower
required concentrations. Each of these options is discussed in the following sections.
Summary of Contents for HAPSITE ER
Page 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Page 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Page 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Page 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Page 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Page 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Page 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Page 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Page 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Page 435: ......
Page 436: ......