13 | Data Review
INFICON
256 / 436
074-471-P1E HAPSITE ER Operating Manual
Remove
Removes the
Background Subtract
triangle.
Select Mass for Integration
Selects either the TIC or a mass fragment
for integration.
Show Integration
Displays the integration on the x-axis.
Show Retention Time
Displays the retention time on the x-axis.
13.8.3 Range Tool
The Range Tool is a red-striped triangle located at the bottom left side of the
chromatogram. It is used to average spectra over a “range” of scans across a given
peak, especially when the analytes are low in concentration. It can also be used to
select a section of a peak or reintegrate peaks. The SI numbers for the selected
compound will increase when using the
Range Tool
. Follow the instructions below to
use the
Range Tool
.
1
Place the cursor on the red-striped triangle, which is the
Range Tool
. Left-click,
hold and drag the triangle to the location where the scans should be averaged.
Summary of Contents for HAPSITE ER
Page 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Page 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Page 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Page 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Page 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Page 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Page 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Page 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Page 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Page 435: ......
Page 436: ......