14 | Tune
INFICON
296 / 436
074-471-P1E HAPSITE ER Operating Manual
Shifts the mass peak left
Increases the peak resolution.
Decreases the peak resolution.
Shifts the mass peak right.
Decreases the
Ion Energy
.
Increases the
Ion Energy
.
Zooms to a single peak scan display
window.
14.4.12 Setting the Full Scan Range
Placing the mouse cursor on the x axis of the full scan window and right-clicking
displays the
Set Full Scan Range
window. This allows a custom scan range to be
entered. The scan ranges of 45 to 300 amu or 1 - 45 amu can also be selected.
The EM voltage will automatically be decreased by 500 volts (default) whenever a
range below mass 45 is scanned.