12 | ER IQ Software
INFICON
168 / 436
074-471-P1E HAPSITE ER Operating Manual
12.3.4.1 Set Access Level
In the
Set Access Level
option there are two user levels which can be set in
ER IQ,
Normal
and
Advanced.
Neither access level has a factory set password.
Normal
level allows users to run samples, view results, and perform basic operations
with HAPSITE ER.
Advanced
allows access to all user operations. This includes all normal user
functions plus method creation and editing, file deletion, changing alarm parameters,
and changing network settings.
To restrict access to advanced functions, an advanced user password can be set.
Once the password is set, it must be entered each time the
ER IQ
program is opened,
or whenever the access level is changed from normal to advanced. See Setting or
Changing the Access Level Password [
169].
12.3.4.1.1 Changing Access Levels
When the
Normal
access level is selected, a prompt will be displayed stating that
some areas of the
ER IQ
will have restricted access. Click
Yes
if continuing is desired.
1
To change the access level, click on the
Tools
menu on the System Setup
Screen.
2
Select
Set Access Level...
Summary of Contents for HAPSITE ER
Page 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Page 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Page 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Page 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Page 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Page 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Page 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Page 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Page 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Page 435: ......
Page 436: ......