sat 805 Pulse Oximeter
Disinfecting, Cleaning and Maintenance
90
10.2
Disinfection
It has been proven, wipe disinfection is more thorough and comprehensive than spray disinfection. We
recommend to use Meliseptol rapid (manufacturer B.Braun) in the same manner like for the above outlined
cleaning procedure.
10.2.1
Disinfecting the Unit
ATTENTION! Danger to the Unit!
Do not use disinfectants other than those authorised. Proceed particularly cautiously in the LCD
display area to avoid scratching the surface. Never immerse the unit in or under water i.e. into
another liquid. The unit as well as the applicable Masimo accessories do not sustain autoclave,
steam or gas sterilization.
1.
Switch off the unit before cleaning and disconnect the mains and patient connections.
2.
The application instructions of the disinfectant manufacturer must be followed. Disinfection takes
place as part of a surface wipe disinfection. For this purpose, the disinfectant is distributed on a
disposable wipe and the surface of the pulse oximeter to be disinfected is wiped off using the cross
method (2 x longitudinal/2 x transverse). After the manufacturer- and disinfectant-dependent
exposure time, further use can be made.
3. Allow the device to dry completely before the next use.
10.2.2
Disinfecting the Masimo sensors
NOTICE!
The single patient sensors (adhesive sensors) are only intended for use on one patient and must not
be cleaned or reused.
ATTENTION: Danger to sensor functionality!
Do not use undiluted bleach 15% - 5.25% sodium hypochlorite or any cleaning solution other than
those recommended here because permanent damage to the sensor may occur.
Do not immerse the sensor or connector in any liquid solution.
Do not sterilize by irradiation, steam autoclave or ethylene oxide!