sat 805 Pulse Oximeter
Alarms
82
Status Line Message
Description
Solution
Ambient Light!
The unit has determined a
scattered or foreign light source.
This can occur with particularly
strong lighting from the outside
(especially from xenon or similar
lamps).
Do not expose the sensor to a
direct light source, or shade from
the outside.
Status messages are not alarms and there not accompanied by LED activation.
Status messages may be mask with higher ranking alarms.
8.4
Battery Alarms
The device warns the user in good time of low or insufficient battery capacity, and reminds the user in good
time to connect the pulse oximeter again to the mains supply and to recharge the internal battery.
There are three stages in battery alarm management:
–
Battery alarm low priority
–
Battery alarm medium priority
–
System battery alarm (high priority)
WARNING! Insufficient Monitoring!
With continued battery alarm act quickly. Monitoring of the patient can discontinue in a short
time.
Reconnect to the device to mains and recharge the unit.
Once connected to the mains power supply, the battery alarm will be automatically reset.
However if the recharging is interrupted early, and the device is again run but not off the mains, the battery
alarm may immediately be triggered again.