sat 805 Pulse Oximeter
Operation
39
The lifetimes are defined as follows:
Sensor / Cable
Active Patient
Monitoring Limit
Duration if
Monitoring 24
Hours Per Day
Duration if
Monitoring 12
Hours Per Day
Duration if
Monitoring 8
Hours Per Day
Single-patient-use
SpO2-Sensors with
replaceable tape
336 hours
14 days
28 days
42 days
Single-patient-use
SpO2-Sensors without
replaceable tape
168 hours
7 days
14 days
21 days
Reusable SpO2-Sensors
(DCI, DCIP, YI, TC-I, TF-I,
and DBI)
8760 hours
12 months
24 months
36 months
Patient Cables
17520 hours
24 months
48 months
72 months
6.6.2.1
90%-Lifetime
or „Near expiration“
-message
Sensor / Cable
Remaining time at 90%
Single-patient-use SpO2-Sensors with
Replaceable tape
ca. 34 hours
Single-patient-use SpO2-Sensors without
Replaceable tape
ca. 17 hours
ReusableSpO2-Sensors
(DCI, DCIP, YI, TC-I, TF-I, and DBI)
ca. 876 hours
Patient Cables
ca. 1752 hours