sat 805 Pulse Oximeter
42
After the cause of the alarm has been eliminated, this alarm is downgraded again to the technical message
"Sensor expired".
If, on the other hand
1.
the sensor-pulse oximeter connection is disconnected or
2.
longer than 2 hours the sensor would not have been on the patient or
3.
a device reset (e.g. by a restart) is performed,
the sensor is immediately switched as defective (alarm: defective sensor) and further use is impossible.
Fig. 30: Defective sensor
High priority alarm
“Defective sensor” (beyond lifetime)
6.7
Using the Interfaces
The interfaces on the device may be operated and installed by trained personnel only. Connecting the
analog or nurse call interface via the interface connector is only allowed upon instruction of these persons.
WARNING! Danger by improper Application of the Interface!
If non-specified devices are connected to the interfaces of the device, there may be health-related
injuries of the patient caused by equalizing or leakage currents.
For assuring a complete patient insulation, connect electrically isolated devices only.