sat 805 Pulse Oximeter
Alarms
81
Low Priority Alarms:
Status Line Message
Description
Solution
Mains power supply!
The mains power or the
DC power supply is
interrupted/defect.
Check the mains power (fuses) and
also the DC power pack. Is the green
LED glowing?
Loudspeaker defective!
The sat 805 has detected a
defective alarm
loudspeaker and has
switched over to the
internal substitute signal
device.
A defective loudspeaker can only be
replaced by qualified personnel. Notify
your medical technology dealer.
8.3
Status message
The following status messages provide information about the actual monitoring situation:
Status Line Message
Description
Solution
Low perfusion!
The unit has determined
insufficient circulation for reliable
determination of oxygen saturation
values.
Remove the sensor and position
at a different place.
Signal IQ too low!
The signal strength for signal
reliability is too low.
Remove the sensor and position
at a different place.
Pulse search!
At too low perfusion or during the
initialization, the device attempts
to synchronize with the detected
perfusion.
If the device will remain in this
state after switching on or this
message is displayed during the
operation, remove the sensor and
position at a different place.
Interference!
The unit has determined influence
from a second light source or a
second sensor.
Remove the interfering source.
Only use a new Masimo sensor.