sat 805 Pulse Oximeter
Control
66
–
Perfusion sensitivity setting
–
Signal averaging time setting
NOTICE!
The displayed information is read-only.
1.
Starting from the main menu, call:
Product Info
2.
Leave that window by confirming
Return
.
Fig. 67: Device Information
7.7.7
Analogue Interface (
Analogue out
)
If required a calibration signal can be output at the analog interface.
For calibration, a null signal (0%), a full signal (100%) or a step function can be output.
–
0 %
: On the interface, the analog voltages for 0% SpO
2
, 0 1/min pulse rate and the baseline for the
plethysmogram are output.
–
100 %
: On the interface, the analog voltages for 100% SpO
2
, 240 1/min pulse rate and the full-scale
deflection of the plethysmogram are output.
–
Steps
: In this mode, voltage steps are output, starting at 0 volts and ending at 1.0 volt, rising by
incremental steps of 0.1 volt per second.
1.
Starting from the main menu, call:
Configuration
Analogue
out
2.
Select the desired function (
0 %
/
100 %
/
steps
).
NOTICE!
The calibration signals are immediately
available on the interface.
3.
Leave the menu by selecting
Return
.
Fig. 68: Analogue Interface
Fig. 69: Calibration