sat 805 Pulse Oximeter
Technical Data
19
3.3
Factory Default Settings
Setting
Value
Alarm tone volume
5
Pulse tone volume
5
Alarm tone mute time
60 sec
Contrast
5
Averaging Time
8 sec
Perfusion sensitivity
normal
Nurse Call
N(ormally) O(pen)
Alarm limit SpO
2
high
100 %
Alarm limit SpO
2
low
85 %
Alarm limit Pulse high
160 1/min
Alarm limit Pulse low
40 1/min
Alarm filter SpO
2
low
OFF
Alarm filter Pulse high
OFF
SmartTone
ON
Artifact Filter
OFF
3.4
Life Times
Component
Value
sat 805
5 years
Batteries
3 years
Patient cable
17280 hours, X-CAL - limited
Pulse oximetry sensor, reusable
8760 hours, X-CAL - limited
Pulse oximetry sensor, not reusable
168 hours, X-CAL - limited
The Masimo X-CAL feature will prevent the use of components with an error message after the
abovementioned hours. A use after this expected life could lead by faulty measurements to a possible risk
for the patient. For further information on Masimo X-CAL, please refer to chapter 6.6.2.