sat 805 Pulse Oximeter
Malfunctions and Troubleshooting
87
Message
Cause
Solution
Status Line Message:
"Adh. Sensor near
expiration!"
The adhesive sensor has reached
90% of its specified life time.
Have a new sensor ready.
Status Line Message:
"X-CAL: Patient cable
expired!"
The patient cable has reached
100% of its specified life time.
Replace patient cable.
Status Line Message:
"XCAL:Sensor expired!"
The sensor has reached 100% of
its specified life time.
Replace sensor.
Status Line Message:
"X-CAL:Adh. sensor
expired
The sensor has reached 100% of
its specified life time.
Replace sensor.
Status Line Message:
"Check sensor
connection!"
The system has detected a
possible fault in the accessories.
Check connection between
patient cable and sensor. If
needed, replace sensor and/or
cable.
Status Line Message:
"Check cable and
sensor!"
The system has detected a
possible fault in the accessories.
Check connection between
patient cable and sensor. If
needed, replace sensor and/or
cable.
Status Line Message:
„No adh. sensor
connected“
The sensor is not or incorrectly
connected to the unit.
Check sensor. If needed, replace
sensor.
Status Line Message:
„No patient cable
connected“
The patient cable is not or
incorrectly connected to the unit.
Check patient cable. If needed,
replace cable.
Status Line Message:
„Unrecognized cable“
The unit has discovered a patient
cable not approved for this
system.
Use approved Masimo
accessories only.