sat 805 Pulse Oximeter
Control
60
NOTICE!
During the alarm silence time, the alarm
silence symbol
appears to indicate
that the option is active (the number
below shows the remaining time in
seconds).
WARNING! New alarms cannot be detected!
If a new alarm condition occurs during the alarm tone mute duration, the new alarm will be
displayed after completion of the mute duration.
7.7.3
Masimo Signal Processing (
Signalprocessing
)
Starting from the main menu, call:
Configuration
Signalprocessing
Following actual signal processing parameters
can be set:
–
Averaging time (
Averaging time
)
–
Sensitivity (
Algorithm mode
)
–
Smart Tone (
ON/OFF
)
–
Artifact-filter (
ON/OFF
)
Fig. 61: Masimo Signal Processing
7.7.3.1 Averaging Time (
Averaging time
)
The averaging time indicates the duration, with which from several original measured values individual
displays one VALUE is again computed.
The averaging time can be adjusted between 2-4 (FastSat
TM
) and 16 seconds. 8 seconds is the default
setting.