sat 805 Pulse Oximeter
Operation
37
WARNING! Allergy Risk!
The use of Masimo and RD SET disposable SpO
2
-sensors is contraindicated on patients who
develop allergic reactions to the tape. The sensors must be re-positioned every 8 hours - and as
soon as lack of blood circulation or skin surface injury is displayed - remove them again and attach
them in another area.
6.6.1.2
Reusable Sensors
p/n
Sensor Type
Patient Type and Weight
Application Site
HPO02-DCSC
DCSC
Children and Adults > 30 kg, Spot
Check-Applications
Ring- or Middle Finger of the
non-dominant Hand
HPO04-DCI
LNCS
®
DCI
Children and Adults > 30 kg
Ring- or Middle Finger of the
non-dominant Hand
HPO04-DCIP
LNCS
®
DCIP
Children and Adults > 10 kg and
< 50 kg
Ring- or Middle Finger of the
non-dominant Hand
HPO04-YI
LNCS
®
YI
Children and Adults > 1 kg
> 10 kg: Ring- or Middle Finger
of the non-dominant Hand.
> 3 kg < 10 kg: Great Toe.
< 3 kg: Foot, alternatively Back
of the Hand.
HPO04-TCI
LNCS
®
TC-I
Adults and Children > 30 kg,
Earsensor
Earlobe or Auricle
HPO04-TFI
LNCS
®
TF-I
Forehead Sensor, transflective,
Adults > 30 kg
Forehead
HPO05-DCI
RD SET DCI
Children or Adults > 30 kg
Ring- or Middle Finger of the
non-dominant Hand
HPO05-DCIP
RD SET DCIP
Children or Adults > 10 kg and
< 50 kg
Ring- or Middle Finger of the
non-dominant Hand