sat 805 Pulse Oximeter
Control
46
1.
Press ON/OFF button (1).
2.
A signal sound activates.
3.
An optical, audio, and internal self
examination is completed:
–
The LEDs are activated.
–
A dark and bright test picture
appears briefly on the LC display.
–
A short alarm signal is issued.
The unit goes through extensive
internal hardware testing.
Fig. 33 Switching on
ATTENTION! Unsuccessful Tests!
If one or several tests are not successfully executed, a corresponding error message is displayed in
the display.
The unit can and may not be used on patients.
ATTENTION! Defective Alarm Loudspeaker!
If the alarm loudspeaker is defective, a replacement signal generator is activated internally. This
does not take over the full functionality of the alarm loudspeaker (volume, tone sequences).
The device must then not be used on the patient!
4.
After this the Normal Display
appears.
Now the signal search for the pulse
occurs (blinking dashes and sensor
symbol). With a successful pulse search
the measured values for oxygen
saturation and pulse frequency are
displayed on the LCD screen.
Fig. 34: Normal Display