sat 805 Pulse Oximeter
Operation
40
The following technical messages are output when 90 % of the specified service life of the patient cable or
sensor is reached.
Fig. 21: Cable near expiration
Technical message:
„Cable near expiration“
Fig. 22: Sensor near expiration
Technical message:
„Sensor near expiration“
Fig. 23: Adh. sensor near expiration
Technical message:
„Adh. sensor near expiration“
6.6.2.2
Life time exceeded
If the life time of the accessories is reached during monitoring operation, the technical message "Service
life reached" is output.
Exceeding this threshold triggers a further period of time: the so-called "extension phase". This is 12 hours
for single-patient adhesive sensors and 72 hours for reusable sensors.
In this phase, the accessories can still be used for the current monitoring, but may be before a switch-off
with a sensor-defective message.
Fig. 24: X-CAL: Cable expired
Technical message:
„X
-
CAL: Cable expired“