sat 805 Pulse Oximeter
Malfunctions and Troubleshooting
84
9
Malfunctions and Troubleshooting
WARNING! Insufficient Monitoring by Malfunctions!
With the occurrence of function failure the patient cannot be correctly monitored in certain cases.
The cause of the function failure must be removed as soon as possible.
Immediately assure monitoring of patient by other means.
Action at Function Failure:
1.
Secure monitoring of patient by other means.
2.
Remove unit from patient.
3.
Remove external interfaces.
4.
Correct failure with help of Table below.
WARNING! Danger to the Patient!
Never use a malfunctioning unit.
List of possible Malfunctions / Error Messages:
Message
Cause
Solution
Unit cannot be turned on.
Batteries are empty.
Recharge batteries: connect unit
to electricity.
Should function not occur even
with a networked connection,
inform medical device supplier.
The information displayed on the
LCD is difficult to read.
The display contrast is set
incorrectly.
Set the display contrast that the
information is easy to read.
When in use buttons do not work. Internal failure.
Inform medical device supplier.