sat 805 Pulse Oximeter
Control
57
4.
Set the correct number.
5.
Repeat steps 3 and 4 until the complete
code is entered.
6.
Confirm the code by selecting
Accept
.
NOTICE!
An invalid access code is answered by
the error message:
"
Incorrect
access code!
“
Re-enter the correct
code.
Fig. 56: Invalid Access Code
7.7.2
Acoustical Alarms (
Acoust. Alarms
)
Starting from the main menu, call:
Configuration
Acoustic. Alarms
Following settings can be adjusted:
–
Pulse tone volume
(
Volume Pulse
)
–
Alarm tone volume (
Volume Alarm
)
–
Alarm tone mute time (
Silence
Time
)
Fig. 57: Acoustical Alarm
7.7.2.1
Pulse Tone Volume (
Volume Pulse
)
When the pulse signal tone is activated, then in monitoring mode a tone signal will be played for every
detected pulse. The pitch of the tone identifies the current oxygen saturation. In other words, the higher
the pitch of the tone, the greater is the measured oxygen saturation (and vice versa).
1.
Starting from the main menu, call:
Configuration
Acoust.
Alarms
Volume Pulse
2.
Adjust the pulse tone volume with the
trim knob from
OFF
to loud (
10
).
Fig. 58: Pulse Tone Volume