sat 805 Pulse Oximeter
Control
47
NOTICE!
In case the Masimo sensor is not
connected or not correctly on the
patient, a corresponding error message
is displayed, e.g.
"Sensor off
Patient!"
.
In addition an alarm sound is heard.
Fig. 35: Sensor off Patient!
NOTICE!
If the alarm tone volume setting is 5 or
less, a warning message is displayed
during the power-up cycle.
The operator has to acknowledge this
setting with OK.
Otherwise, the device will not enter the
standard mode but will be switched off
automatically.
5
.
Now the unit is operational.
Fig. 36: Silent Alarm Tone!
7.4
Turning the Unit off
The turn-off process is designed so that turning it off accidentally can possibly be avoided. After turning off
the unit all previously entered values and configurations are preserved.
Except for the settings
„
Alarm tone loudness
= 0 (OFF)“ and „algorithm“ the minimum audible =1 and
=“Norm.Perfus.“ is returned.
To turn the unit OFF:
1.
Press the ON/OFF-button.
2.
The power-down menu is displayed in
the LC display.
3.
With the trim knob, select
Turn off
and confirm.
Fig. 37: Power-Down Menu