sat 805 Pulse Oximeter
Alarms
83
8.4.1
Battery Alarm (low priority)
Status Line Message
Description
Solution
Battery low!
Remaining operation time of
one hour or less.
Recharge battery.
A flashing battery icon will be displayed.
8.4.2
Battery Alarm (medium priority)
Status Line Message
Description
Solution
Battery depleted!
Remaining operation time of
15 minutes or less.
Recharge battery.
A flashing battery icon will be displayed.
8.4.3
Battery Alarm (high priority)
Status Line Message
Description
Solution
SYST ALARM: Bat
low!
Battery capacity is so low,
that reliable monitoring can
no longer be guaranteed.
Signal processing and the
monitoring function are
disabled.
Recharge battery.
An empty battery icon will be displayed.
8.5
Combination of Alarms of Different Priorities
With overlapping different alarm conditions the latest alarm with the highest priority is activated.
With concurrently occurring alarms the highest ranking is activated. If these alarm conditions are removed,
the next lower waiting alarm is immediately displayed.
A new alarm condition of higher priority immediately supersedes any lower ranking alarm.