Table 40 Warning event codes (continued)
Details and Solution
Description
Event code
NOTE:
The PIN is required every time a token
is installed in the USB port and every time the
library is powered on.
Parity error
58
•
Retry the operation.
Error log overflow
59
•
If the problem continues, contact Hewlett
Packard Enterprise customer support.
Disable encryption before downgrading firmware.
Unable to downgrade firmware while encryption
enabled.
5A
Invalid cartridge. LTO3 media is not supported
with encryption enabled.
5B
•
Verify that the tape cartridge is supported.
•
Update the tape drive firmware to the latest
version to use LTO3 media.
Change the network settings to IPv4 before
downgrading firmware.
IPv6 will not be supported after downgrading
firmware.
5C
Drive firmware is specific to the drive model and
interface. Select the drive firmware specific to this
Wrong drive firmware — invalid drive firmware
type.
5D
drive. If the firmware file could have been
renamed, obtain a new copy of the firmware file
In an MSL4048, a full-height tape drive cannot
be installed in the middle two drive bays; it must
Illegal drive position of FH drive.
5E
be installed in either the upper two drive bays or
lower two drive bays. Reinstall the tape drive in
a supported drive bay.
The device firmware cannot be downgraded in
cases that would compromise encrypted data,
No firmware downgrade possible.
5F
critical component support, or license terms.
Check the configuration settings. If the device
has been using the Encryption Kit, you might need
to disable encryption before downgrading
firmware. If the Library Extender is installed, you
might need to uninstall the Library Extender
before downgrading firmware.
Complete the cleaning process and retry the
operation.
Cleaning tape installed.
60
Cleaning failure. Cleaning process could not be
performed.
61
•
Check cleaning tape and exchange if
necessary.
•
Retry the operation.
Exchange the cleaning tape.
Cleaning tape expired.
62
Invalid cartridge. Drive has rejected the cartridge
as invalid.
63
•
Verify that the cartridge generation and
technology are supported by the device. See
“Tape cartridges” (page 43)
.
Invalid cleaning cartridge. Drive has rejected the
cartridge as invalid.
64
•
Retry the operation.
•
If problems persist, exchange the data
cartridge.
Warning events
177