Caution
Do not short power-supply voltages to ground. The analyzer power-supply
current limiting cannot protect the resistors in series with the power supply.
Note
Some transistors have collectors connected to the case. Electrical connection
of the case to the collector might not be reliable, making collector voltage
measurements on the transistor case unreliable.
IF Signature
1. Disconnect W27 (coax 3) from
2. Connect the source connection of a 3
power splitter to
(Minicircuits Model: ZSC
J-2-l) Connect one output of the power splitter to the input of an HP
Spectrum
Analyzer. Connect the other output of the power splitter to cable W27 (coax 3).
Note
If a 3
power splitter is not available, an SMB tee and an active probe with
a
divider may be substituted. Connect the active probe between the tee
and the other spectrum analyzer. The absolute power levels are approximately
17
lower than those stated below, due to the elimination of the 3
power
splitter and the 20
loss through the
divider.
IF Section 8-23
Summary of Contents for 8560A
Page 93: ......
Page 150: ... ...
Page 178: ...4 20 Replaceable Parts ...
Page 180: ...4 22 Replaceable Parts ...
Page 182: ...4 24 Replaceable Parts ...
Page 184: ...4 26 Replaceable Parts ...
Page 186: ...4 28 Replaceable Parts ...
Page 188: ...4 30 Replaceable Parts ...
Page 190: ...4 32 Replaceable Parts ...
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Page 249: ...6 44 General Troubleshooting ...
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 282: ......
Page 307: ......
Page 319: ......
Page 324: ...8 42 IF Section ...
Page 325: ......
Page 326: ......
Page 327: ......
Page 344: ......
Page 397: ......
Page 398: ......
Page 423: ......
Page 441: ...12 18 Display Power Supply Section ...
Page 442: ......
Page 443: ......
Page 449: ...A 6 Component Level Information Packets ...
Page 450: ......