307
ACC
308 FREQ ACC
309
ACC
310
ACC
311
ACC
312
ACC
313 FREQ ACC
314 FREQ ACC
Transfer oscillator pretuned DAC out of range. The transfer oscillator
pretune procedure attempts to find pretuned DAC values by
programming the PLL to 25 different frequencies and incrementing
the transfer oscillator pretune DAC until XFRSENSE changes
polarity. ERR 307 is set if the DAC is set to 255 (maximum) before
XFRSENSE changes polarity.
Offset oscillator pretune DAC not within prescribed limits at low
frequency. The offset oscillator pretune DAC is set to provide a
frequency less than 189 MHz while the PLL is programmed for 189
MHz. ERR 308 is set if XFRSENSE is greater than
V (it should
be at the negative rail).
Offset oscillator pretune DAC not within prescribed limits at high
frequency. The offset oscillator pretune DAC is set to provide a
frequency less than
MHz while the PLL is programmed for 204
MHz. ERR 309 is set if XFRSENSE is greater than
V (it should
be at the negative rail).
Main roller pretune DAC set to 255. The main roller pretune DAC is
set to 5, causing MAINSENSE to go to the positive rail. The DAC
is incremented until MAINSENSE changes polarity. ERR 310 is set
if the DAC is set to 255 before MAINSENSE changes to a negative
polarity.
Main roller pretune DAC set to 255. The main roller pretune DAC is
set to 5, causing MAINSENSE to go to the positive rail. The DAC
is incremented until MAINSENSE changes polarity. ERR 311 is set
if the DAC is set to 255 before MAINSENSE changes to a negative
polarity.
Unable to adjust MAINSENSE to 0 volts using the fine adjust DAC.
The coarse adjust and fine adjust DAC are used together to set
MAINSENSE to 0 volts with the loop opened. ERR 312 is set if the
fine adjust DAC cannot bring MAINSENSE to 0 volts.
Error in LO synthesis algorithm. ERR 313 is set if a combination of
Sampler oscillator and roller oscillator frequencies could not be found
to correspond to the required YTO start frequency. Contact the
factory.
Indicates problems in the span calibration. Troubleshoot any unlocks
before attempting to troubleshoot span calibration problems, because
the loops must all lock in order to perform the calibration. If LO
spans greater than 1 MHz are correct, check
,
or
This error message appears when the main
roller oscillator sweep sensitivity is 0. A sweep ramp is injected into
the locked main roller loop which should generate a negative-going
ramp on MAINSENSE. ERR 314 is set if the slope of this ramp is 0.
This is an indication of an unlocked main roller loop or lack of a sweep
ramp.
General Troubleshooting 6-19
Summary of Contents for 8560A
Page 93: ......
Page 150: ... ...
Page 178: ...4 20 Replaceable Parts ...
Page 180: ...4 22 Replaceable Parts ...
Page 182: ...4 24 Replaceable Parts ...
Page 184: ...4 26 Replaceable Parts ...
Page 186: ...4 28 Replaceable Parts ...
Page 188: ...4 30 Replaceable Parts ...
Page 190: ...4 32 Replaceable Parts ...
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Page 249: ...6 44 General Troubleshooting ...
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 282: ......
Page 307: ......
Page 319: ......
Page 324: ...8 42 IF Section ...
Page 325: ......
Page 326: ......
Page 327: ......
Page 344: ......
Page 397: ......
Page 398: ......
Page 423: ......
Page 441: ...12 18 Display Power Supply Section ...
Page 442: ......
Page 443: ......
Page 449: ...A 6 Component Level Information Packets ...
Page 450: ......