17. Signal ID Oscillator Adjustment
9. While observing the HP
display, adjust
SIG ID for the highest
obtainable frequency with less than 3
decrease in amplitude from maximum. Read this
frequency from the frequency counter and record as
HIGH =
Hz
10. Observe the HP
display as you adjust
SIG ID for the lowest obtainable
frequency with less than 3
decrease in amplitude from maximum. Record the
frequency counter reading as
L O W
=
MHz
11. Calculate the difference between
and
then divide results by four.
Enter the result as
=
12. Add
to
recorded in step 10 and record the result as
F
=
MHz
13. Adjust
for a frequency counter reading equaling
The final adjusted
frequency must equal 10.7 MHz
Adjustment Procedures 2-65
Summary of Contents for 8560A
Page 93: ......
Page 150: ... ...
Page 178: ...4 20 Replaceable Parts ...
Page 180: ...4 22 Replaceable Parts ...
Page 182: ...4 24 Replaceable Parts ...
Page 184: ...4 26 Replaceable Parts ...
Page 186: ...4 28 Replaceable Parts ...
Page 188: ...4 30 Replaceable Parts ...
Page 190: ...4 32 Replaceable Parts ...
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Page 249: ...6 44 General Troubleshooting ...
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 282: ......
Page 307: ......
Page 319: ......
Page 324: ...8 42 IF Section ...
Page 325: ......
Page 326: ......
Page 327: ......
Page 344: ......
Page 397: ......
Page 398: ......
Page 423: ......
Page 441: ...12 18 Display Power Supply Section ...
Page 442: ......
Page 443: ......
Page 449: ...A 6 Component Level Information Packets ...
Page 450: ......