A3 Assembly’s Video Circuits
Voltages from
to
Variable Gain Amplifier correspond (approximately) to
on-screen signal levels. (One volt corresponds to the top of the screen and zero volts
corresponds to the bottom of the screen.) This is true for both log and linear settings except
when the analyzer is in 1 dB/div or 2 dB/div. In these cases the log expand amplifier is
selected, and 1 V corresponds to top-screen and 0.8 or 0.9 V corresponds to bottom-screen.
The analyzer can be set to zero span at the peak of a signal to generate a constant dc voltage
in the video circuits during sweeps.
1. Disconnect W26 from
and
from
2. Connect W26 to
3. Set the HP 8560A to the following settings:
SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SWEEP TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 ms
RES BW
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
LOG/div
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..lOdB/div
4. If a trace is displayed, troubleshoot the A3 assembly. If a trace is absent, connect an
oscilloscope to the rear-panel BLANKING OUTPUT.
5. The presence of a TTL signal (TTL low during 20 ms sweep) indicates a good A3
Interface Assembly. Troubleshoot the IF Section.
6. If the BLANKING OUTPUT is always at a TTL high or low, troubleshoot the
trigger circuits.
7. Reconnect W26 to
and
to
8. Remove the A3 assembly’s shield.
9. If the video filters appear to be faulty, refer to “Video Filter” in this chapter.
10. If there appears to be a peak detector problem, refer to “Positive/Negative Peak
Detectors” in this chapter.
11. Connect the HP 8560A’s CAL OUTPUT to the
follows:
INPUT
and set the controls as
300 MHz
0 Hz
REFLVL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
C E N T E R F R E Q
. . . . . . . . . . . . . . . . . . . . . . . .
SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12. If the analyzer works correctly in 5 dB/div and 10 dB/div but not in 1
or
2
refer to “Log Offset/Log Expand” in this chapter. Continue with step 13 if the
problem involves on-screen amplitude errors which appear to originate in the video chain.
13. Monitor
with an oscilloscope. If the voltage is not approximately
Vdc,
troubleshoot the A4 Log Amplifier. (Refer to the IF troubleshooting procedure.)
14. To confirm proper video input to the video circuit, change the reference level in 10
steps from -10
to
At each 10
step, the input voltage should change
100
The input level should be
Vdc for a
reference level.
7-12
Section
Summary of Contents for 8560A
Page 93: ......
Page 150: ... ...
Page 178: ...4 20 Replaceable Parts ...
Page 180: ...4 22 Replaceable Parts ...
Page 182: ...4 24 Replaceable Parts ...
Page 184: ...4 26 Replaceable Parts ...
Page 186: ...4 28 Replaceable Parts ...
Page 188: ...4 30 Replaceable Parts ...
Page 190: ...4 32 Replaceable Parts ...
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Page 249: ...6 44 General Troubleshooting ...
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 282: ......
Page 307: ......
Page 319: ......
Page 324: ...8 42 IF Section ...
Page 325: ......
Page 326: ......
Page 327: ......
Page 344: ......
Page 397: ......
Page 398: ......
Page 423: ......
Page 441: ...12 18 Display Power Supply Section ...
Page 442: ......
Page 443: ......
Page 449: ...A 6 Component Level Information Packets ...
Page 450: ......