Procedure 1. Analyzer Cover
Procedure 1. Analyzer Cover
Removal/Replacement
1. Disconnect the line-power cord, and place the analyzer on its front panel.
2. Loosen (but do not remove) the four rear-bumper screws, using a 4 mm hex wrench. Pull
the cover assembly off towards the rear of the instrument.
Caution
When replacing the analyzer’s cover, use caution to avoid damaging any
cables.
3.
4.
When installing the cover assembly, be sure to locate the cover’s air vent holes on the
bottom side of the analyzer. Attach with the four screws loosened in step 2, and tighten
the four screws gradually to ensure that the cover is seated in the front-frame gasket
groove.
Torque each screw to 40 to 50 inch-pounds to ensure proper EM1 gasket compression.
Figure 3-1. Hinged Assemblies
Assembly Replacement 3-3
Summary of Contents for 8560A
Page 93: ......
Page 150: ... ...
Page 178: ...4 20 Replaceable Parts ...
Page 180: ...4 22 Replaceable Parts ...
Page 182: ...4 24 Replaceable Parts ...
Page 184: ...4 26 Replaceable Parts ...
Page 186: ...4 28 Replaceable Parts ...
Page 188: ...4 30 Replaceable Parts ...
Page 190: ...4 32 Replaceable Parts ...
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Page 249: ...6 44 General Troubleshooting ...
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 282: ......
Page 307: ......
Page 319: ......
Page 324: ...8 42 IF Section ...
Page 325: ......
Page 326: ......
Page 327: ......
Page 344: ......
Page 397: ......
Page 398: ......
Page 423: ......
Page 441: ...12 18 Display Power Supply Section ...
Page 442: ......
Page 443: ......
Page 449: ...A 6 Component Level Information Packets ...
Page 450: ......