Procedure 2. Al Front Frame/Al8 CRT
Figure 3-5. Installing the CRT and Front-Frame Assemblies
Replacement
Note
Use care when handling the glass CRT EM1 shield. The glass may be cleaned
using thin-film cleaner (HP part number 8500-2163) and a lint-free cloth.
When installing the glass shield, face the side of the glass with the silver
coated edge towards the inside of the analyzer.
1. Place the analyzer on its right-side frame with the front end extending slightly over the
front of the work bench.
2. Gently place the A18 CRT assembly into the Al Front-Frame assembly as llustrated in
Figure 3-6.
3. Place the front-frame and CRT assemblies into the analyzer, using caution to avoid
pinching any cables.
4. Dress the
trace-align wires between the CRT assembly mounts and the A6 Power
Supply top shield.
3-8 Assembly Replacement
Summary of Contents for 8560A
Page 93: ......
Page 150: ... ...
Page 178: ...4 20 Replaceable Parts ...
Page 180: ...4 22 Replaceable Parts ...
Page 182: ...4 24 Replaceable Parts ...
Page 184: ...4 26 Replaceable Parts ...
Page 186: ...4 28 Replaceable Parts ...
Page 188: ...4 30 Replaceable Parts ...
Page 190: ...4 32 Replaceable Parts ...
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Page 249: ...6 44 General Troubleshooting ...
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 282: ......
Page 307: ......
Page 319: ......
Page 324: ...8 42 IF Section ...
Page 325: ......
Page 326: ......
Page 327: ......
Page 344: ......
Page 397: ......
Page 398: ......
Page 423: ......
Page 441: ...12 18 Display Power Supply Section ...
Page 442: ......
Page 443: ......
Page 449: ...A 6 Component Level Information Packets ...
Page 450: ......