Procedure 6. A6 Power Supply Assembly
16. Use a screwdriver to remove three standoffs from the A6 Power Supply assembly.
17. Remove the A6 Power Supply assembly by lifting from the regulator
toward front
of analyzer.
Replacement
1. Attach the A6 Power Supply assembly to the analyzer’s chassis using the three standoff
screws.
2. Connect
to
W3 to
fan power wires to
to
and the
line-power jack to
See Figure 3-11.
3. Secure the
High Voltage assembly to the A6 Power Supply assembly, using three
screws and washers. Connect ribbon cable
to
4. Snap post-accelerator cable
to the CRT assembly.
F A N
POWER
L I NE
POWER
W I
Figure
A6 Power Supply Connections
5. Ensure that all cables are safely routed and will not be damaged when securing the A6
cover.
6. Secure the power supply cover shield to the power supply using three
screws (1).
See Figure 3-12. One end of the cover fits into a slot provided in the rear frame assembly.
3-20 Assembly Replacement
Summary of Contents for 8560A
Page 93: ......
Page 150: ... ...
Page 178: ...4 20 Replaceable Parts ...
Page 180: ...4 22 Replaceable Parts ...
Page 182: ...4 24 Replaceable Parts ...
Page 184: ...4 26 Replaceable Parts ...
Page 186: ...4 28 Replaceable Parts ...
Page 188: ...4 30 Replaceable Parts ...
Page 190: ...4 32 Replaceable Parts ...
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Page 249: ...6 44 General Troubleshooting ...
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 282: ......
Page 307: ......
Page 319: ......
Page 324: ...8 42 IF Section ...
Page 325: ......
Page 326: ......
Page 327: ......
Page 344: ......
Page 397: ......
Page 398: ......
Page 423: ......
Page 441: ...12 18 Display Power Supply Section ...
Page 442: ......
Page 443: ......
Page 449: ...A 6 Component Level Information Packets ...
Page 450: ......