3. Check for excessive LO feedthrough. Use the LO Feedthrough performance test in the
Installation and Verification Manual, but check over a Center Frequency range of 300
to 100 MHz. The LO Feedthrough will be 3.9107
greater than the Center Frequency
setting.
Harmonic/Spurious Outputs Too High
Harmonic and spurious outputs may be generated by A10 itself or may be present on the
either the 600 MHz drive or 1st LO drive signal. There is a direct relationship between
spurious signals on the 1st LO and spurious signals on the TG output. There is a five-to-one
relationship between spurious signals on the 600 MHz drive and the spurious signals on the
TG output. For example, if the 600 MHz signal moves 1 MHz, the TG output signal will
move 5 MHz. This is due to the multiplication in the pentupler.
1. If the Harmonic Spurious Responses performance test failed, connect another spectrum
analyzer, such as an HP
to the HP 8560A
LO OUTPUT connector. Set the
HP 8560A to each frequency as indicated in the performance test, with the Span set to
0 Hz. At each frequency
MORE OF 2, FREQ DIAGNOSE,
LO FREQ
. The frequency displayed will be the fundamental frequency of the
LO
OUTPUT. Use the HP
to measure the level of the second and third harmonics of
the 1st LO signal.
Note
The 1st LO typically has a higher harmonic content than the tracking
generator output. For the purposes of this check, it is the variation in
harmonic content versus frequency which is important.
If the variation of the harmonic level of the 1st LO versus frequency tracks the harmonic
level variation of the tracking generator output, repeat this step while measuring the 1st
LO signal at the end of W43 nearest
If there is little variation in the 1st LO harmonic
level between the
LO OUTPUT connector and W43, and the relative variation in
harmonic level tracks with the TG output harmonic level, suspect either the A7 LODA or
A11 YTO.
If the harmonic level variation of the 1st LO versus frequency does not track the harmonic
level variation of the TG output, suspect
2. If sidebands are present at the same frequency offset at every output frequency, check
the spectral purity of the 1st LO and the 600 MHz drive signals using another spectrum
analyzer. When checking the 1st LO, the HP 8560A must be set to zero span. Press
MORE OF 2, FREQ DIAGNOSE, LO FREQ
to determine the 1st LO frequency. A 1 MHz
sideband on the 1st LO will appear as a 1 MHz sideband on the output signal.
To verify that the 600 MHz drive or 1st LO signal is responsible for the sidebands,
substitute a clean signal for the 600 MHz drive or 1st LO signal. If the sidebands on the
output disappear when using the clean signal, the substituted signal was responsible for the
sidebands.
Note
The 600 MHz drive signal should be -8
+/-3.5
The 1st LO signal
should be
11-22 RF Section
Summary of Contents for 8560A
Page 93: ......
Page 150: ... ...
Page 178: ...4 20 Replaceable Parts ...
Page 180: ...4 22 Replaceable Parts ...
Page 182: ...4 24 Replaceable Parts ...
Page 184: ...4 26 Replaceable Parts ...
Page 186: ...4 28 Replaceable Parts ...
Page 188: ...4 30 Replaceable Parts ...
Page 190: ...4 32 Replaceable Parts ...
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Page 249: ...6 44 General Troubleshooting ...
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 282: ......
Page 307: ......
Page 319: ......
Page 324: ...8 42 IF Section ...
Page 325: ......
Page 326: ......
Page 327: ......
Page 344: ......
Page 397: ......
Page 398: ......
Page 423: ......
Page 441: ...12 18 Display Power Supply Section ...
Page 442: ......
Page 443: ......
Page 449: ...A 6 Component Level Information Packets ...
Page 450: ......