YTO Leveling Loop (334)
This error often requires troubleshooting the Al4 Frequency Control
assembly or A7 LODA (Synthesizer Section) or the ADC circuits.
334 LO AMPL
LO Distribution Amplifier is unleveled. This error is usually
accompanied by error codes 300 or 301. ERR 301 YTO UNLK is
cleared once ERR 334 has been cleared. Check the output of the
All YTO with the jumper on
in the TEST position. The
YTO power output should be between
and
If the
YTO is working properly, refer to “A7 LODA Drive” in Chapter
10. The LODA AGC voltage is monitored by the ADC. ERR 334 is
set if LODA AGC is outside of its prescribed limits. Refer to “A7
LODA Drive” in Chapter 11. Error 334 may also be displayed if the
front-panel LO OUTPUT is not terminated into 50 ohms.
Sampling Oscillator (335)
This error indicates an unlocked Sampling oscillator (also known as
the offset Lock Loop).
335 SMP UNLK
Sampling oscillator PLL is unlocked. ERR 335 is set if OFL_ERR is
outside its prescribed limits.
Automatic IF Errors (400 to 599)
These error codes are generated when the automatic IF adjustment
routine detects a fault. This routine first adjusts amplitude
parameters, then resolution bandwidths in this sequence: 300
1
MHz, 100
30
10
3
1
300 Hz, 100 Hz, 30 Hz,
10 Hz, and 2 MHz. The routine restarts from the beginning if a fault is
detected. Parameters adjusted after the routine begins but before the
fault is detected are correct; parameters adjusted later in the sequence
are suspect. Refer to “Automatic IF Adjustment” in Chapter 8.
The IF Section relies on the ADC and video circuitry to perform its
continuous IF adjustments. IF-related errors occur if the ADC, video
circuitry, or A4 assembly linear path is faulty.
400 AMPL 100
401 AMPL 300
402 AMPL
403 AMPL 3K
404 AMPL
Unable to adjust amplitude of 100 Hz Resolution Bandwidth.
Unable to adjust amplitude of 300 Hz Resolution Bandwidth.
Unable to adjust amplitude of 1
Resolution Bandwidth.
Unable to adjust amplitude of 3
Resolution Bandwidth.
Unable to adjust amplitude of 10
Resolution Bandwidth.
Errors 405 to 416:
When these
resolution bandwidth (RBW) error
messages appear, use the following steps to check for errors 581 or 582.
1. Set the spectrum analyzer (LINE) switch OFF.
2. Set the spectrum analyzer
switch ON and observe the lower
right-hand corner of the display for 10 seconds.
6-22 General Troubleshooting
Summary of Contents for 8560A
Page 93: ......
Page 150: ... ...
Page 178: ...4 20 Replaceable Parts ...
Page 180: ...4 22 Replaceable Parts ...
Page 182: ...4 24 Replaceable Parts ...
Page 184: ...4 26 Replaceable Parts ...
Page 186: ...4 28 Replaceable Parts ...
Page 188: ...4 30 Replaceable Parts ...
Page 190: ...4 32 Replaceable Parts ...
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Page 249: ...6 44 General Troubleshooting ...
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 282: ......
Page 307: ......
Page 319: ......
Page 324: ...8 42 IF Section ...
Page 325: ......
Page 326: ......
Page 327: ......
Page 344: ......
Page 397: ......
Page 398: ......
Page 423: ......
Page 441: ...12 18 Display Power Supply Section ...
Page 442: ......
Page 443: ......
Page 449: ...A 6 Component Level Information Packets ...
Page 450: ......