6. Turn off the analyzer and place Al7 in the service position. Connect the ground lead of
a high-voltage probe (HP 34111A) to the chassis, and use it with a DVM to measure
10).
7. The nominal
voltage is -1600 Vdc, but the CRT will function with this voltage
within 200 V of -1600 Vdc.
8. Adjusting
COURSE FOCUS should vary the
voltage by 150 V. If
these voltages are correct, suspect the CRT.
9. Check the
High Voltage Module cathode supply output at
using a high
voltage probe. If the cathode voltage is -2450 V +/-250 V, check the Focus Grid Level
Shifter.
10. If the cathode voltage is not correct, check the
High Voltage Module and its
connections.
11. Connect an oscilloscope probe to
This signal corrects the focus for the X
position of the CRT beam, and for intensity level. It also provides the front-panel focus
adjusting voltage.
12. Press
MORE, and FOCUS. While turning the front-panel knob, verify the dc level
of the signal at
adjusts about 30 Vp-p.
13. Verify that the front panel
adjustment plus
Z FOCUS changes the
peak-to-peak voltage at
by 25 V.
14. Set front panel
to minimum and Z FOCUS fully counterclockwise and
X
FOCUS to fully counterclockwise. Verify that the peak-to-peak voltage at
is about
40 Vp-p (due to X-Dynamic Focus circuit).
15. If circuit operation seems correct,
CRT is probably at fault.
Intensity Problems
Intensity problems, or absence of display, can be due to the Al7 assembly’s Intensity
Amplifier (function block A), Intensity Grid Level Shifter (function block B), CRT
interconnections, or lack of proper supplies or inputs to A17.
1. On the HP 8560A press
MORE 1 of 2 ,
.
2. Rotate the front-panel knob (RPG), and check that the intensity changes from dim, but
readable, to bright.
3. If the
function does not function properly, troubleshoot the A2 Controller
assembly.
4. Perform the preliminary and Z-axis portions of the Display Adjustment in Chapter 2.
Verify that
CUTOFF functions properly. If
CUTOFF does not function
properly, place the Al7 CRT Driver in the service position.
5. Verify that blanking pulses are present at
using an oscilloscope. The pulses
should be normal TTL levels, approximately 1
wide and 4 or 7
apart. If the
blanking pulses are not correct, check the BLANKING output of the A2 Controller
Assembly and cable W7.
Display/Power Supply Section 12-9
Summary of Contents for 8560A
Page 93: ......
Page 150: ... ...
Page 178: ...4 20 Replaceable Parts ...
Page 180: ...4 22 Replaceable Parts ...
Page 182: ...4 24 Replaceable Parts ...
Page 184: ...4 26 Replaceable Parts ...
Page 186: ...4 28 Replaceable Parts ...
Page 188: ...4 30 Replaceable Parts ...
Page 190: ...4 32 Replaceable Parts ...
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Page 249: ...6 44 General Troubleshooting ...
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 282: ......
Page 307: ......
Page 319: ......
Page 324: ...8 42 IF Section ...
Page 325: ......
Page 326: ......
Page 327: ......
Page 344: ......
Page 397: ......
Page 398: ......
Page 423: ......
Page 441: ...12 18 Display Power Supply Section ...
Page 442: ......
Page 443: ......
Page 449: ...A 6 Component Level Information Packets ...
Page 450: ......