2. Display Adjustment
Fast Zero Span Adjustments
22. Set
SWEEP OFFSET,
VIDEO OFFSET,
VIDEO GAIN and
SWEEP GAIN to midrange. Adjustment locations are shown in Figure 2-4 for
these A2 adjustments.
23. Set the HP 355D Attenuator to provide 30
attenuation.
24. Press
on the analyzer, and connect the equipment as shown in Figure 2-2. Set
the HP 8560A controls as follows:
C E N T E R F R E Q
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...300 MHz
SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 Hz
REF LVL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
RES BW
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
VIDEO BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sweep time
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 ms
25. Press
MARKER-> REF LVL .
If the marker is not at the top graticule, press
MARKER-B REF LVL
again.
Press
(SAVE), SAVE STATE,
and
STATE 0.
27. Set sweeptime to 10 ms.
28.
Press
(SAVE), SAVE STATE,
and
STATE 1 .
29. Adjust
SWEEP OFFSET to place the beginning of the trace at the leftmost
vertical graticule line.
30. Adjust
SWEEP GAIN to place the end of the trace at the tenth vertical graticule
line (one division from the right edge of the graticule).
31. Press
and press the (up) key seven times.
Press
SAVE STATE
, and
STATE 2.
Set the sweep time to
50 ms. Press
SAVE STATE,
and
STATE 3 .
Press
RECALL STATE,
and
STATE 1 .
Switch between
STATE1
and
STATE2.
Adjusting
and
so that in STATE1
the trace is lined up with the top graticule and in STATE2 the trace is lined up with the
eighth graticule (from the top line). Repeat until they align to within
divisions.
36. Adjust
and
until the start of sweep is aligned to the leftmost vertical
graticule line and the stop sweep is aligned with the right most vertical graticule line.
37. Press STATE2
and
STATE3.
The two traces should be aligned /-.l divisions.
Press STATE0
and
STATE1
. The two traces should be aligned /-.l divisions.
Adjustment Procedures 2-19
Summary of Contents for 8560A
Page 93: ......
Page 150: ... ...
Page 178: ...4 20 Replaceable Parts ...
Page 180: ...4 22 Replaceable Parts ...
Page 182: ...4 24 Replaceable Parts ...
Page 184: ...4 26 Replaceable Parts ...
Page 186: ...4 28 Replaceable Parts ...
Page 188: ...4 30 Replaceable Parts ...
Page 190: ...4 32 Replaceable Parts ...
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Page 249: ...6 44 General Troubleshooting ...
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 282: ......
Page 307: ......
Page 319: ......
Page 324: ...8 42 IF Section ...
Page 325: ......
Page 326: ......
Page 327: ......
Page 344: ......
Page 397: ......
Page 398: ......
Page 423: ......
Page 441: ...12 18 Display Power Supply Section ...
Page 442: ......
Page 443: ......
Page 449: ...A 6 Component Level Information Packets ...
Page 450: ......