4. IF Amplitude Adjustments
A5 Reference Attenuator Adjustment
12. Set the HP 3335A (AMPLITUDE) to -60
13. Connect a BNC cable between the
50
output of the HP 3335A and the HP 8560A
INPUT
14. On the HP
press (CAL) and
REF LEVEL ADJUST.
Use the front-panel knob or step
keys to place the peak of the displayed signal 3
to 5
below the reference level.
On the HP 8560A press
[
PEAK SEARCH
]
and
MARKER DELTA.
Set the analyzer reference
level to -10
16. Change the HP 3335A
to -10
17. On the HP 8560A press
18. Note the AMKR amplitude. Ideally, it should read 50.00
+/-.l
19. If the AMKR amplitude is less than 49.9
rotate
(15
one-half
turn counterclockwise for each 0.1
below 50.00
If the AMKR amplitude is greater
than 50.1
rotate
one-half turn clockwise for each 0.1
above 50.00
Do
not adjust
more than five turns before continuing with the next step.
20. On the HP 8560A press
ADJ
IF STATE.
Note the AMKR amplitude reading.
Set the HP 8560A reference level to -60
and press
and
MARKERS OFF .
22. Repeat steps 12 through 21 until the AMKR amplitude reading is 50.00
+/-.l
A5 Adjustment Verification
23. On the HP
disconnect W29 from
Connect the test cable between
and
the 50 output of the HP 3335A.
24. Set the HP 8560A reference level to -10
25. Set the HP 3335A
to -5
On the HP 8560A press
and
MARKER NORMAL.
27. The MARKER amplitude should read -10
+/-.13
If the reading is outside of
this range, repeat steps 4 through 21.
28. On the HP
reconnect W29 to
Press
and set the controls as follows:
C E N T E R F R E Q
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 Hz
REF LVL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
RES BW
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300kHz
29. Connect a BNC cable between the HP 8560A CAL OUTPUT and INPUT
30. On the HP
press
(CAL) and
REF LVL ADJ .
31. Use the knob or step keys to adjust the REF LEVEL CAL setting until the MKR reads
-10.00
+/-.l
32.
On the HP
press
STORE REF LVL .
2-28 Adjustment Procedures
Summary of Contents for 8560A
Page 93: ......
Page 150: ... ...
Page 178: ...4 20 Replaceable Parts ...
Page 180: ...4 22 Replaceable Parts ...
Page 182: ...4 24 Replaceable Parts ...
Page 184: ...4 26 Replaceable Parts ...
Page 186: ...4 28 Replaceable Parts ...
Page 188: ...4 30 Replaceable Parts ...
Page 190: ...4 32 Replaceable Parts ...
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Page 249: ...6 44 General Troubleshooting ...
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 282: ......
Page 307: ......
Page 319: ......
Page 324: ...8 42 IF Section ...
Page 325: ......
Page 326: ......
Page 327: ......
Page 344: ......
Page 397: ......
Page 398: ......
Page 423: ......
Page 441: ...12 18 Display Power Supply Section ...
Page 442: ......
Page 443: ......
Page 449: ...A 6 Component Level Information Packets ...
Page 450: ......