Adjustments
Model 3585A
5-48
i. Set the external attenuators for 32dB of attenuation.
j . Set the 3585A REFERENCE LEVEL to -60dBm.
k. Adjust A1 8R7 1 for an offset marker amplitude readilng of -32.00dBm.
1.
Set the external attenuators for 48dB of attenuation.
m. Set the 3585A REFERENCE LEVEL to -76dBm.
n. Adjust A1 8R65 for an offset marker amplitude reading of -48.00dB.
o. Disconnect the Tracking Generator from A17J 1 . Reconnect the cable from A6CJ 1 to
A17J 1 .
Summary of Contents for 3585A
Page 264: ......