Model 3585A
Adjustments
5·2. Equipment Required
Table 5-2 lists the equipment required to perform the adjustments on the 3585A. Equipment
that meets or exceeds the required characteristics given in the table may be substituted for
the recommended models.
Table 5·2. Recommended Adjustment Equipment
Equipment
Required Characteristics
Recommended Model
Digital Volt/Ohmmeter
DC Volts: 2V, 20V, 200V range
-hp- 3466A
Accuracy: ± 0.04%
I nput I mpedance 1 0MO
Ohms: 200
Accuracy: ± 0 . 0 7 %
H i g h Frequency A C
A C Volts: 0 . 3V,3V range
-hp- 400E
Voltmeter
Frequency Response: 1 00Hz to
1 M Hz
I nput I mpedance: 1 0MO
Accuracy: ± 1 %
Oscilloscope
Bandwidth : dc to 1 00MHz
-hp- 1 7 40A
Vertical Range: 5mV/div to 20V/div
Horizontal Range: 50nsec/div to
1 00msec/div
H igh Voltage Probe
Accuracy:
::s;
1 %
-hp- 3440A- K05
I nput I mpedance:
�
1 09 Ohms
Measurement Range:
�
6 kV
Frequency Counter
Frequency Range: 1 0Hz to 1 50MHz -hp- 5382A
Accuracy: ± 1 count ± time base
error
Resolution: 0. 1 Hz
Frequency Synthesizer
Frequency Range: 200Hz to 40MHz -hp- 3 3 3 5A
Amplitude Accuracy: ± 0 . 2 7 d B
Attenuator 1 dB/Step
Attenuation Range: 0 to 1 2dB
-hp- 3 5 5 C
Accuracy: ± 0 . 2 dB
Frequency Range: dc to 40MHz
Attenuator 1 0dB/Step
Attenuator Range: 0 to 1 00dB
-hp- 3 5 5 D
Accuracy: ± 0 . 5dB
Frequency Range: dc to 40MHz
9 M Hz Low Pass Filter
( See Figure 4-2 1 )
1 OkO Resistor
± 1 % , 1 /8 Watt
-hp- P . N . 0 7 5 7-0442
BNC-To-Sealectro
Supplied with instrument
-hp- P. N . 0 3 5 8 5-6 1 6 1 6
Adapter Cable
Optional
Spectrum Analyzer
Frequency Range: 1 00kHz to
-hp- 8 5 5 8 B
1 50MHz
Amplitude Accuracy: ± 3dB
Resistor Probe
20: 1 Resistive Divider
-hp- 1 0020A
1 kO
Resistance
5-3
Summary of Contents for 3585A
Page 264: ......