Performance Tests
Model 3585A
4-12
k. Wait until the sweep is completed. The trace you now see is the flatness of the 750
input.
1.
Repeat Steps d thru f and enter the results on the Operational Verification Test Card.
4· 1 6. 1 Ohm Frequency Response Test
This test checks the frequency response of the 1 MO input relative to the flatness of the 500
input. Ideally the difference between the two signals would be zero.
Specification:
High Impedance Frequency Response
20Hz
10MHz
40MHz
Equipment Required:
500 Feed Thru Termination . . . . . . . . . -hp- 1 l048C
Procedure:
a. Set the 3585A controls for:
INSTRUMENT PRESET
STOP FREQUENCY . . . . . . . . . . . . . . . . . . 1O MHz
dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l dB
RANGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 dBm
RES. BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 kHz
RES. BW HOLD . . . . . . . . . . . . . . . . . . . . . . . . . . on
RECALL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
b . Connect the 3585A Tracking Generator output to the 3585A 500 input.
c. Adjust the Tracking Generator Amplitude control for the cent er of its range.
d. Allow one complete sweep to occur. Press the STORE A -- B key of the 3585A.
e. Connect a 500 termination to the 1 MO input of the 3585A. Connect the output of the
Tracking Generator to the input of this termination.
f. Set the 3585A controls for:
INPUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
l
MO
B TRACE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . off
A-B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
g. Move the marker to the most negative point on the displayed trace. (Ignore the LO
feedthrough point at 0 Hz)