Operation Overview
Model 3585A
3-22
REMJTE
o
REMOTE
o
TALK
o
SRO
o
L I STEN
o
Hp·IB STATUS FU NCTI O N S *
�.§D
Lights to indicate that the 3585A is in the Remote control mode . This mode can be
entered only via the HP-lB.
(When the 3585A is in Remote, all front-panel functions except the LINE
switch , the DISPLAY controls , the Tracking Generator AMPLITUDE
control and the LOCAL key are disabled . Pressing any key
(except
or
�m
) or rotating the Continuous Entry control
will cause the beeper to sound and the message, "HP-IB REMOTE SET "
to appear on the CRT screen .).
Returns the 3585A to Local and reenables all front-panel functions . An HP-IB
Local Lockout will disable
until a remote Return To Local command is
given , or the LINE switch is turned off and then back on .
L I STEN
o
TALK
o
(Pressing
during an HP-IB Local Lockout causes the beeper to
sound and the message, "HP-IB LOCAL LOCKOUT" to appear on the
CRT screen .)
Lights to indicate that the 3585A is addressed to listen . **
Lights to indicate that 3585A is addressed to talk .
Lights to indicate that the 3585A is generating an HP-IB Service Request .
*HP-IB operation is fully described in the Operating Manual .
* *The LISTEN or TALK light will remain on (even in Local) until the 3585A is unaddressed via the HP-IB or is
turned off and then back on.